Recall: GlucoMen LX Sensor blood glucose test strips

$
0
0

More than 1.7 million blood glucose testing strips used by diabetics are being recalled over potential false readings.

GlucoMen LX Sensor blood glucose test strips 50 count (product code 38877), used with the GlucoMen LX and GlucoMen LX PLUS meter manufactured by Menarini.

  • Lot 3212214249, expiry date 31-08-2014
  • Lot 3212219249, expiry date 31-08-2014.

Affected lots have been contaminated during manufacture, and may give an overestimation of blood glucose levels, potentially resulting in inappropriate insulin administration.

The manufacturer distributed the affected lots in the UK via wholesale outlets during October and November 2012. The product may, however, still remain with users or pharmacies.

If people have the faulty test strips at home they can telephone the manufacturer’s Menarini on 0800 243667 who will provide free replacements. If they have questions, they can contact their GP, diabetes clinic or another healthcare professional who can give them advice about what alternative test strips they can use.

There is currently no evidence that UK patients have been affected.